| Product Model | TRED30-16 (Replaceable Battery) |
| External Sensor Measurement Range | -40°C to +99°C (-40°F to +210°F) |
| Operating Temperature Range | -30°C to +60°C (-22°F to +140°F) |
| Storage Temperature Range | -10°C to +40°C (14°F to +104°F) |
| Temperature Accuracy | -20°C to +40°C: ±0.5°C (typ. ±0.3°C) -40°C to -20°C: ±0.8°C (typ. ±0.5°C) +40°C to +70°C: ±0.8°C (typ. ±0.7°C) +70°C to +99°C: ±1.2°C (typ. ±1.0°C) Actual performance typically exceeds rated values; recalibration can improve accuracy |
| Temperature Resolution | -40°C to +50°C: 0.1°C +50°C to +70°C: 0.2°C +70°C to +80°C: 0.3°C +80°C to +99°C: 0.4°C – 0.6°C LogTag Analyzer displays 0.1°C resolution; internal resolution stored in device |
| Sensor Response Time | Typically < 2 minutes (T90, 1 m/s airflow) |
| Memory Capacity | 15,905 temperature readings |
| Logging Duration | Up to 66 days (6-minute interval) |
| Daily Summary (LCD) | Up to 30 days (Max/Min/Duration and alarm events) |
| Sampling Interval | Configurable from 30 seconds to several hours |
| Start Options | Push-button start or date/time start, delay up to 18 hours |
| Recording Indicator | “RECORDING” status indicator |
| Download Time | Typically < 10 seconds for full memory (15,905 readings) |
| Ingress Protection | IP61 (vertical or hanging installation) |
| Power Supply | 3V CR2032 LiMnO₂ coin cell (user replaceable) |
| Battery Life | Up to 1 year storage + 1 year operation under normal use (6-minute logging, daily LCD check ≤30s, monthly download)Note: frequent alarms reduce battery life |
| Real-Time Clock | Built-in Accuracy: ±25 ppm (~2.5 sec/day) Temp coefficient: -0.034 ±0.006 ppm/°C |
| Probe Cable Length | Standard: 1.5 m Max recommended: 3 m |
| Probe Cable Type | PTFE (FDA food-contact compliant) |
| Dimensions | 93 mm (H) × 54.5 mm (W) × 8.6 mm (T) |
| Weight | 41 g |
| Housing Material | Polycarbonate |